Measurement item
|
Total Organic Carbon
|
Measurement principle
|
Continuous Wet Oxidation + NDIR detector
|
Measuring range
|
Rance A:0-2000ppb Range B:0-30.0ppm(possible measure up + 050.0ppm)
|
Measuring period
|
Continuous measurement
|
Reaction Time
|
Approximately 5 min. (response rate 90%)
|
Water pressure
|
0-0.5MPa(Automatic aspiration)
|
Sample flow rate
|
> 100ml/min
|
Sample Temp
|
5-50¡É
|
Resolution
|
Range A:5ppb Range B:0.1ppm
|
Repetitive accuracy
|
Range A: ±2%FS Range B: ±2%FS
|
Reagent
|
Phosphoric acid for both range A and B
|
Oxidant
|
Sodium persulfate for only range B
|
Calibrating Solution
|
Potassium Hydrogen phthalate
|
Career gas
|
Highly purified Nitrogen gas
|
Alarm output
|
High and low TOC value, instrument disorder,
2 points relay contact, AC 125V/0.5A
|
Signal output
|
2-20mA TOC measurement value Max. burden 600Mohm
|
Setting environment
|
Ambient temperature 5-40¡É Ambient humidity 5-80%
|
Dimensions/Weight
|
1160(H)×480(W)×400(D) min
Approx.60kg(Incl.mobile mount and umbo excl. tires)
|
Power supply
|
Adjustable to foreign power supply 50/60Hz (Adjustable to foreign power supply)
|